Features
- Reading / display of registered parameters
- Measured system (phase and wires)
- Energy / power calibration function
- Energy / power reference output (pulse)
- Conversion time and synchronization
- Latch up function for instant reading
- Communication via remote control (RS-232)
Model & Ordering
Model : HC 3100 A protable energy testing devices (with wide current input range 0 ~ 100A)
Option features:
- HD 108 photo pick up sensor (Together with a built-in 5V power supply needs to quote separately).
- 10A Voltage / current test leads(Totally 10pcs)
- 100A test leads(Totally 4 sets)
Specification
- Accuracy of Input Ranges :
- Voltage Input :
- Range selection : Fully auto-ranging
- Vab / Vbc / Vca : 0 ~ 600V Line to line voltage, input
/ 1.2M
- Van / Vbn / Vcn : 0 ~ 440V Line to neutral voltage, input
/ 600K
- Resolution : 0.001V
- Accuracy : 0.05% RD for 50 ~ 600V range; 0.05% RG for 6 ~ 600V range
- Current Input :
- Range selection : Fully auto-ranging
- Ia / Ib / Ic : A10 0 ~ 10A range / 0 ~ 10A input terminals, 2VA / phase; A100 0 ~ 100A range / 0 ~ 100A input terminals, 2VA / phase
- Resolution : 0.00001A / 0.0001A (A10), 0.0001A / 0.001A (A100)
- Accuracy 0.05% RD, 0.25~10A for A10 range, 5~100A for A100 range; 0.05% RG, 0~0.25A for A100 range, 0~5A for A100 range
- Power / Energy :
- Range selection : Fully auto-ranging
- Power factor : Watts/watthours for cosØ ; vars/varhours for sinØ
- Accuracy :
- Power Factor :
- Range : 0~1.0000 (leading to lagging)
- Accuracy : 0.05% RG
- Resolution : 0.0001
- Phase Angle :
- Range : 0.0~359.9o
- Accuracy < 0.03o between 1~359o
- Resolution : 0.001o
- Frequency :
- Range : 45~65 Hz
- Accuracy : 0.02% RD
- Resolution : 0.0001 Hz
- Energy Calibration Function :
- Using HC 3100 as a Calibration Reference :
- Accuracy : 0.05% RD (refer to energy / power input ranges)
- Error detection : % error in reading
- Measured input : Digital signals from energy meters Pulse or open collector
- Logic of input : Level high > 1.8V, level low < 0.8V
- Input range : 0.001 WH to 99.999 WH per pulse or 30~300000 pulses per kWh
- Test interval (N) : Numbers of input test pulses, 1~99
- Using HC 3100 Pulse Output for External Comparison :
- Application : HC 3100 digital output to comparator and other reference standards
- Digital output : Refer to HC 3100 digital pulse output
¡@ |