Portable Power / Energy Testing Device HC 3100

Features
  • Reading / display of registered parameters
  • Measured system (phase and wires)
  • Energy / power calibration function
  • Energy / power reference output (pulse)
  • Conversion time and synchronization
  • Latch up function for instant reading
  • Communication via remote control (RS-232)
Model & Ordering

      Model : HC 3100 A protable energy testing devices (with wide current input range 0 ~ 100A)

         Option features:
  • HD 108 photo pick up sensor (Together with a built-in 5V power supply needs to quote separately).
  • 10A Voltage / current test leads(Totally 10pcs)
  • 100A test leads(Totally 4 sets)

 

Specification
  • Accuracy of Input Ranges :
    • Voltage Input :
      • Range selection : Fully auto-ranging
      • Vab / Vbc / Vca : 0 ~ 600V Line to line voltage, input / 1.2M
      • Van / Vbn / Vcn : 0 ~ 440V Line to neutral voltage, input / 600K
      • Resolution : 0.001V
      • Accuracy : 0.05% RD for 50 ~ 600V range; 0.05% RG for 6 ~ 600V range
    • Current Input :
      • Range selection : Fully auto-ranging
      • Ia / Ib / Ic : A10 0 ~ 10A range / 0 ~ 10A input terminals, 2VA / phase; A100 0 ~ 100A range / 0 ~ 100A input terminals, 2VA / phase
      • Resolution : 0.00001A / 0.0001A (A10), 0.0001A / 0.001A (A100)
      • Accuracy 0.05% RD, 0.25~10A for A10 range, 5~100A for A100 range; 0.05% RG, 0~0.25A for A100 range, 0~5A for A100 range
    • Power / Energy :
      • Range selection : Fully auto-ranging
      • Power factor : Watts/watthours for cosØ ; vars/varhours for sinØ
    • Accuracy :
    •       
    • Power Factor :
      • Range : 0~1.0000 (leading to lagging)
      • Accuracy : 0.05% RG
      • Resolution : 0.0001
    • Phase Angle :
      • Range : 0.0~359.9o
      • Accuracy < 0.03o between 1~359o
      • Resolution : 0.001o
    • Frequency :
      • Range : 45~65 Hz
      • Accuracy : 0.02% RD
      • Resolution : 0.0001 Hz
  • Energy Calibration Function :
    • Using HC 3100 as a Calibration Reference :
      • Accuracy : 0.05% RD (refer to energy / power input ranges)
      • Error detection : % error in reading
      • Measured input : Digital signals from energy meters Pulse or open collector
      • Logic of input : Level high > 1.8V, level low < 0.8V
      • Input range : 0.001 WH to 99.999 WH per pulse or 30~300000 pulses per kWh
      • Test interval (N) : Numbers of input test pulses, 1~99
      •       
    • Using HC 3100 Pulse Output for External Comparison :
      • Application : HC 3100 digital output to comparator and other reference standards
      • Digital output : Refer to HC 3100 digital pulse output
¡@