HC PP series transmitters are a family of plugging unit type, set in a standard DIN 19" rack mounting or in an independent unit-enclosure which according to user application. Each plugging unit is designed & divided into two parts : input terminal module & converting module.
PPT-1 is the plug-in type Thermocouple Isolating transducer with one output.
Measuring and conversion of IEC 688 . The adaptability for power system of VDE 0110, and Measuring reliability of VDE 3540.
PPT-1 is the plug-in type Thermocouple Isolating transducer with one output.
Measuring and conversion of IEC 688 . The adaptability for power system of VDE 0110, and Measuring reliability of VDE 3540.
- Linear output
- Field – rangeability with switchable input & output ranges
- Changeable input T/C module of J K E T S R B N type
- Outstanding CJC traceability < 0.5℃ maximum 10 minutes warm up
- Accuracy 0.3%
- High dielectric strength 4KV AC rms input / output / power / case
- Available standard or custom calibration
- Accuracy ( 23 ± 5 ℃ ) : 0.3% RO
- Linearity & repeatability :
- 0.12% typical
- Linearized output vs input temperature
- Temperature stability : Less 0.01% fs per degree C
- Configuration :
- Field rangeability
- Changeable thermocouple input module
- Selectable output for voltage or current mode
- Switchable input & output ranges
- Switchable AC power for 110V or 220V
- Selectable C / F calibration
- CJC stability : Less 0.03℃ per ℃, 10 to 40℃
- CJC trace ability :
- Warm up condition, typically
- Less 0.5℃ deviation in 10 minutes
- Input resistance : Minimum 20MΩ
- Effect of wire resistance : Less 4 uV per 100Ω
- Input break detection :
- Selectable Hi-set or Lo-set
- Hi-set over 110% of rated output
- Lo-set less 0.5% of rated output
- Maximum input over : Maximum 250V rms 30 sec
- Power effect : Less 0.003% for per volt change
- Response time & ripple : Less 400ms typically, ripple less 0.15% rms of span
- Output drive capability : Maximum 10mA for voltage mode, maximum 10V for current mode
- Common mode rejection : > 100db 50 / 60Hz
- Dielectric strength :
- 4KV AC between input / output / power / case
- Impulse & surge test : IEEE 587 / 1983 or IEC 255-3
- Storage condition : Temperature range -25 to +70℃, RH 20 to 95% non condensed
- Operating condition : Temperature range -20 to +65℃, RH 20 to 95% non condensed
- Power supply : +20% of rating, less 5VA
- Dimension : W*H*D : 12T (60.6mm) x 3U (128.5mm) x 238mm